Two-Wire 4-20mA Two-Electrode Ozone Sensor For Drinking Water, Direct Drinking Water, Service Process Water
Two-electrode system ozone sensor OZ1.2, used in drinking water, direct drinking water, service and process water, measuring water without surfactant, optional range 0~0.5/2/5/10/20ppm
The functional characteristics of the ampere film ozone sensor include high sensitivity and fast response, which can accurately measure the ozone concentration in water. Its advantages are strong anti-interference ability, suitable for various water quality environments, and long service life. It is suitable for drinking water and industrial water treatment to ensure water quality safety.

Specifications
|   |  			 			 
 
 OZ1.2  |  		
| Indicator |   Ozone 			 |  		
| Application |  			 Swimming pool water, drinking water, service water, process water  |  		
| Measuring system | Membrane covered, amperometric 2-electrode system | 
|  			 
 
 
 
 Electronic  |  			 			 Analog version: - voltage output - not galvanically isolated electronics - analog internal data processing - output signal: analog (analog-out/analog) Digital version: - electronic is completely galvanically isolated - digital internal data processing - output signal: analog (analog-out/digital) or digital (digital-out/digital) mA-version: - current output analog - not galvanically isolated electronics - output signal: analog (analog-out/analog)  |  		
|  			 
 Information about the measuring range  |  			 			 The actual slope of a sensor can vary production-related between 65% and 150% of the nominal slope Note: With a slope > 100% the measuring range is reduced accordingly. (Ex.: 150% slope → 67% of the specified measuring range)  |  		
|  			 Slope drift At repeatability conditions (25 °C, pH 7,2 in drinking water)  |  			Approx. <- 1% per month | 
|  			 
 Working temperature  |  			 			 Measuring water temperature: 0 … +45 °C (no ice crystals in measuring water)  |  		
| Ambient temperature: 0 … +55 °C | |
|  			 Temperature compensation  |  			 			 Automatically, by an integrated temperature sensor  |  		
|  			 
 
 Max. allowed working pressure  |  			 			 Operation without retaining ring: - 0.5 bar - no pressure impulses and/or vibrations  |  		
|  			 Operation with retaining ring in Flow cell: - 1.0 bar, - no pressure impulses and/or vibrations (see option 1)  |  		
|  			 
 
 OZ1.2  |  		||||
|  			 Flow rate (Incoming flow velocity)  |  			Approx. 15-30 l/h (33 – 66 cm/s) in flow FLC, small flow rate dependence is given | |||
| pH-range | pH 1 – pH 11 | |||
| Run-in time | First start-up approx. 1 h | |||
| Response time | T90 : approx. 15 sec. | |||
| Zero point adjustment | Not necessary | |||
| Calibration | At the device, by analytical determination | |||
| Interferences |  			 Cl2: factor 0.03  |  		|||
| Absence of the disinfectant | Max. 24 h | |||
|  			 
 Connection  |  			 			 mV version: Modbus version: 4-20 mA version:  |  			 			 5-pole M12, plug-on flange 5-pole M12, plug-on flange 2-pole terminal or 5-pole M12, plug-on flange  |  		||
| max. length of sensor cable (depending on internal signal processing) | analog | < 30 m | ||
| digital |  			 > 30 m are permissible Maximum cable length depends on application  |  		|||
| Protection type |  			 5-pole M12 plug-on flange: IP68 2-pole terminal with mA-hood: IP65  |  		|||
| Material | Semipermeable membrane, PVC-U, ABS | |||
|  			 
 
 Size  |  			 			 Diameter: Length: mV version Modbus version 4-20 mA version  |  			 			 approx. 25 mm approx. 190 mm (analog signal processing) approx.. 205 mm (digital signal processing) approx. 205 mm approx. 220 mm (2-pole-terminal) approx. 190 mm (5-pole-M12)  |  		||
| Transport | +5 … +50 °C (sensor, electrolyte, membrane cap) | |||
|  			 
 
 OZ1.2  |  		|
|  			 
 
 Storage  |  			Sensor: dry and without electrolyte no limit at +5 … +40 °C | 
|  			 Electrolyte: in original bottle protected from sunlight at +5 … +35 °C min. 1 year or until specified EXP-Date  |  		|
|  			 Membrane cap: in original packing no limit at +5 … +40 °C (used membrane caps can not be stored)  |  		|
|  			 
 Maintenance  |  			 			 Regularly control of the measuring signal, min. once a week The following specifications depend on the water quality: Change of the membrane cap: once a year (depending on the water quality) Change of the electrolyte: every 3 - 6 months  |  		
|  			 Option 1: Retaining ring  |  			 			 
 
 
 
 - When operating with pressures >0.5 bar in flow cell - Dimensions retaining ring 29 x 23.4 x 2.5 mm, slitted, PETP - Different positions for groove selectable(on request)  |  			 			 
 
  |  			![]()  |  		
Technical Data
1. OZ1.2 (analog output, analog internal signal processing)
 A potential-free electrical connection is necessary as the sensor electronic is not equipped with a galvanical
 isolation.
|          |  			 			 Measuring range in ppm  |  			 			 Resolution in ppm  |  			 			 Output Output resistance  |  			 			 Nominal slope in mV/ppm  |  			 			 Voltage supply  |  			Connection | 
| OZ1.2H-M12 | 0.005…2.000 | 0.001 |  			 
 
 0…-2000 mV 1 kΩ  |  			- 1000 |  			 
 
 ±5 - ±15 VDC 10 mA  |  			 			 5-pole M12 plug-on flange 
 Function of wires: PIN1: measuring signal PIN2: +U PIN3: -U PIN4: signal GND PIN5: n. c.  |  		
| OZ1.2N-M12 | 0.05…20.00 | 0.01 | - 100 | |||
| OZ1.2HUp-M12 | 0.005…2.000 | 0.001 |  			 
 
 0…+2000 mV 1 kΩ  |  			+1000 |  			 
 
 10 - 30 VDC 10 mA  |  			 			 5-pole M12 plug-on flange Function of wires: 
 PIN1: measuring signal PIN2: +U PIN3: power GND PIN4: signal GND PIN5: n. c.  |  		
| OZ1.2NUp-M12 | 0.05…20.00 | 0.01 | +100 | 
(Subject to technical changes!)
2.  OZ1.2 (analog output, digital signal processing)
 Analog-out / digital
 -The power supply is galvanically isolated inside of the sensor.
 -The output signal is galvanically isolated too, that means potential-free.
|           |  			 			 Measuring range in ppm  |  			 			 Resolution in ppm  |  			 			 Output Output resistance  |  			 			 Nominal Slope in mV/ppm  |  			Power supply | Connection | 
| OZ1.2H-An-M12 | 0.005…2.000 | 0.001 |  			 analog 0…-2 V (max. -2.5 V) 1 kΩ  |  			- 1000 |  			 
 
 9-30 VDC approx. 20-56 mA  |  			 			 5-pole M12 plug-on flange 
 Function of wires: PIN1: measuring signal PIN2: +U PIN3: power GND PIN4: signal GND PIN5: n. c.  |  		
| OZ1.2N-An-M12 | 0.05…20.00 | 0.01 | - 100 | |||
| OZ1.2H-Ap-M12 | 0.005…2.000 | 0.001 |  			 analog 0…+2 V (max. +2.5 V) 1 kΩ  |  			+1000 | ||
| OZ1.2N-Ap-M12 | 0.05…20.00 | 0.01 | +100 | 
(Subject to technical changes!)
3. OZ1.2 (digital output, digital signal processing)
 -The power supply is galvanically isolated inside of the sensor.
 -The output signal is galvanically isolated too, that means potential-free.
|  			 Measuring range in ppm  |  			 			 Resolution in ppm  |  			 			 Output Output resistance  |  			Power supply | Connection | |
| OZ1.2H-M0c | 0.005…2.000 | 0.001 |  			 Modbus RTU 
 There are no terminating resistors in the sensor.  |  			 			 9-30 VDC 
 approx. 20-56 mA  |  			 			 5-pole M12 plug-on flange Function of wires: PIN1: reserved PIN2: +U PIN3: power GND PIN4: RS485B PIN5: RS485A  |  		
| OZ1.2N-M0c | 0.05…20.00 | 0.01 | 
(Subject to technical changes!)
4. OZ1.2 4-20 mA (analog output, analog internal signal processing)
 A potential-free electrical connection is necessary as the sensor electronic is not equipped with a galvanical
 isolation.
 4.1 Electrical connection: 2 pole terminal clamp
|   |  			 			 Measuring range in ppm  |  			 			 Resolutio n in ppm  |  			 			 Output Output resistance  |  			 			 Nominal slope in mA/ppm  |  			 			 Voltage supply  |  			Connection | 
| OZ1.2MA0.5 | 0.005…0.500 | 0.001 |  			 4…20 mA uncalibrated  |  			32.0 |  			 
 12…30 VDC RL 50Ω…RL 900Ω  |  			 			 
 2-pole terminal (2 x 1 mm²) 
 Recommended: Round cable 4 mm 2 x 0.34 mm²  |  		
| OZ1.2MA2 | 0.005…2.000 | 0.001 | 8.0 | |||
| OZ1.2MA5 | 0.05…5.00 | 0.01 | 3.2 | |||
| OZ1.2MA10 | 0.05…10.00 | 0.01 | 1.6 | |||
| OZ1.2MA20 | 0.05…20.00 | 0.01 | 0.8 | 
(Subject to technical changes!)
4.2 Electrical connection: 5 pole M12 plug-on flange
![]()  |  			 			 Measuring range in ppm  |  			 			 Resolution in ppm  |  			 			 Output Output resistance  |  			 			 Nominal slope in mA/ppm  |  			 			 Voltage supply  |  			Connection | 
| OZ1.2MA0.5-M12 | 0.005…0.500 | 0.001 |  			 4…20 mA uncalibrated  |  			32.0 |  			 12…30 VDC RL 50Ω…RL 900Ω  |  			 			 5-pole M12 plug- on flange 
 Function of wires: PIN1: n. c. PIN2: +U PIN3: -U PIN4: n c. PIN5: n. c.  |  		
| OZ1.2MA2-M12 | 0.005…2.000 | 0.001 | 8.0 | |||
| OZ1.2MA5-M12 | 0.05…5.00 | 0.01 | 3.2 | |||
| OZ1.2MA10-M12 | 0.05…10.00 | 0.01 | 1.6 | |||
| OZ1.2MA20-M12 | 0.05…20.00 | 0.01 | 0.8 | 
(Subject to technical changes!)
Spare Parts
| Type | Membrane cap | Electrolyte | Emery | O-ring | 
| For all OZ1.2 |  			 M20.2  |  			EOZ1/W, 100 ml Art. no. 11101  |  			 			 S1  |  			 			 14 x 1.8 silicone  |  		
(Subject to technical changes!)